| Model Number |
DR8076 |
| Medical Device Problem Codes |
Break (1069); Inflation Problem (1310); Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lot number for the malfunction was provided and a lot history review was performed.Received one dorado pta catheter.The catheter was kinked throughout.A complete circumferential break was noted at the butt joint of the balloon and catheter.Based upon the available information, the definitive root cause is unknown the device is labeled for single use.
|
| |
|
Event or Problem Description
|
|
This report summarizes one malfunction.A review of the reported information indicated that model dr8076 pta balloon dilatation catheter was allegedly difficult to inflate and allegedly leaked and broke.This information was received from a single source.The alleged malfunction involved a patient with no known impact to the patient.The patient was reported to be a (b)(6) year-old male, weighing (b)(6).
|
| |
|