• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT LABORATORIES ARCHITECT BNP CALIBRATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT LABORATORIES ARCHITECT BNP CALIBRATORS Back to Search Results
Catalog Number 08K28-02
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Correction/removal number: 3002809144-06/4/19-006-r.A product recall letter was issued to all architect bnp customers who have received calibrator or control lots still within dating.The letter informs the customer of the issue regarding a time dependent stability issue of the architect bnp calibrators and controls that may lead to controls out of range and shift in control and patient results.The letter informs the customer all architect bnp calibrators and controls will have shortened expiration dates of 165 days from the date of manufacture.The letter instructs the customer to discontinue use of the lots which are beyond the 165 day expiration and destroy any remaining inventory.The cause of the shift is instability of the architect bnp calibrators and controls.
 
Event or Problem Description
The customer observed non-abbott and abbott controls out of range high while using the architect bnp calibrator lot 44k82618.There was no report of incorrect patient results or impact to patient management.
 
Additional Manufacturer Narrative
Further information received from the customer indicated that they observed a shift in non-abbott controls and abbott controls were not affected.The event description was updated to correct this information.No further updates or corrections were necessary.
 
Event or Problem Description
The customer observed non-abbott controls out of range high while using the architect bnp calibrator lot 44k82618.There was no report of incorrect patient results or impact to patient management.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT BNP CALIBRATORS
Common Device Name
BNP CALIBRATORS
Manufacturer (Section D)
ABBOTT LABORATORIES
100 abbott park road
abbott park IL 60064 3500
MDR Report Key8811926
Report Number1415939-2019-00160
Device Sequence Number2657647
Product Code JIT
UDI-Device Identifier00380740017460
UDI-Public00380740017460
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 08/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/15/2019
Device Catalogue Number08K28-02
Device Lot Number44K82618
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/03/2019
Supplement Date Received by Manufacturer07/23/2019
Initial Report FDA Received Date07/22/2019
Supplement Report FDA Received Date08/16/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
3002809144-06/4/19-006R
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARCHITECT I2000SR, LIST 03M74-02; ARCHITECT I2000SR, LIST 03M74-02; SERIAL (B)(4); SERIAL (B)(4)
-
-