• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OBERDORF SYNTHES PRODUKTIONS GMBH 4.3MM DRILL BIT LENGTH 413MM; BIT MILLING STERILE AND NON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OBERDORF SYNTHES PRODUKTIONS GMBH 4.3MM DRILL BIT LENGTH 413MM; BIT MILLING STERILE AND NON Back to Search Results
Catalog Number 03.168.011
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2019, the patient underwent an open reduction internal fixation (orif) surgery for the femoral neck fracture with femoral neck system (fns).During the procedure, while the surgeon was drilling for the locking screw using the drill bit, it took a longer time than usual.The surgeon heard a friction sound and checked the devices and x-rays, but there were no problems such as loosening of devices.It took more than five (5) minutes to drill until contralateral bone, whereas it takes less than a minute normally.During the drilling, the surgeon pulled out the drill bit two (2) times and removed soft tissues which adhered to the drill bit.The color of the soft tissues had turned black-red.No metal powder was found in the soft tissues.The surgeon mentioned that it didn¿t seem that the drill bit interfered with any metal devices.The surgeon could insert the locking screw and completed the surgery without other problems.There were less than thirty (30) minutes of surgical delay.The patient outcome is unknown.Concomitant device reported: unknown locking screw (part# unknown, lot# unknown, quantity# 1).This report is for one (1) 4.3 mm drill bit length 413 mm.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The drill bit was received with the reported condition of not functioning.Visual inspection confirmed that the drill bit is quite dull, the cutting edges are nicked and rounded.The instrument is in used condition.The complaint condition is confirmed as the drill bit is quite dull.The drill bit was manufactured in june 2017 according to the specifications.After a visual inspection it is determined that the reusable instrument (part# 03.168.011) is worn from repeated use and servicing; therefore, further investigation for the reported complaint device is not required.During the investigation, no product design or manufacturing issues were observed that may have contributed to the complaint condition; therefore, further corrective and/or preventive action is not required.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.The root cause was identified during the performed cq evaluation and therefore the in the investigation flow listed remaining investigation steps are not required.Device history lot part: 03.168.011, lot: f-22047, manufacturing site: selzach, supplier: (b)(4), release to warehouse date: 09.June 2017.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.3MM DRILL BIT LENGTH 413MM
Common Device Name
BIT MILLING STERILE AND NON
Manufacturer (Section D)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key8813490
Report Number8030965-2019-66488
Device Sequence Number2768735
Product Code GFG
UDI-Device Identifier07611819657373
UDI-Public(01)07611819657373
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.168.011
Device Lot NumberF-22047
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2019
Initial Date Received by Manufacturer 06/26/2019
Supplement Date Received by Manufacturer07/25/2019
Initial Report FDA Received Date07/22/2019
Supplement Report FDA Received Date08/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - SCREWS: LOCKING; UNK - SCREWS: LOCKING
-
-