| Model Number |
7015 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/04/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed multiple kinks along the hypotube.The hypotube is separated 24.1cm from the hub.Microscopic examination revealed damage to the tip.There is contrast present in the inflation lumen and balloon.There is blood present in the guidewire lumen and the balloon is loosely folded.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 28jun2019.This device was received with no reported issues.However, upon review of this device, the hypotube was separated.
|
| |
|