• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG. 7MM X 23MM STRYKER BIOSTEON SCREW; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG. 7MM X 23MM STRYKER BIOSTEON SCREW; SCREW, FIXATION, BONE Back to Search Results
Model Number 234-010-161
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 06/26/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device manufacture date is not known at this time.Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that the screw broke and part of screw was left inside the patient.
 
Additional Manufacturer Narrative
The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: screw broke.According to biocomposites: "investigation findings are: minimal information has been provided relating to the screw breakage complaint.Warnings and precautions, including prevention of screw breakage are stated in the instructions for use.No medical intervention was required.Surgery was completed successfully.No delay in surgery.If the tapped depth is not deep enough, the bottom of the screw stays stationary and the peak (top half) of the screw continues to twist causing breakage.The suspected root causes are user error in method used- if the tapped depth is not deep enough, the bottom of the screw stays stationary and the peak (top half) of the screw continues to twist causing breakage." manufacture date is not known.
 
Event or Problem Description
It was reported that the screw broke and part of screw was left inside the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG. 7MM X 23MM STRYKER BIOSTEON SCREW
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key8814078
Report Number0002936485-2019-00303
Device Sequence Number3027199
Product Code HWC
UDI-Device Identifier15060155710249
UDI-Public15060155710249
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K003641
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number234-010-161
Device Catalogue Number234-010-161
Device Lot Number0917PH298165
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2019
Initial Date Received by Manufacturer 06/26/2019
Supplement Date Received by Manufacturer06/26/2019
Initial Report FDA Received Date07/22/2019
Supplement Report FDA Received Date02/19/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-