Model Number PM3562 |
Device Problems
Pacemaker Found in Back-Up Mode (1440); Communication or Transmission Problem (2896); No Pacing (3268)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/05/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
The results of the investigation are inconclusive since the device was not returned for analysis.
Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
It was reported that the patient presented on (b)(6) 2019 for implant procedure.
Upon review, the pacemaker lost telemetry with the wand and went into backup vvi.
Also, the pacemaker was intermittently pacing.
The patient was pacer dependent.
The pacemaker was not used and was replaced.
There were no patient consequences reported.
|
|
Manufacturer Narrative
|
The reported field event of backup mode was confirmed in the laboratory due to review of device image.
Device image analysis indicated data corruptions occurred that left the device in a wrong state.
After resetting the power, device regained normal operation and backup mode was no longer reproduced.
Original device image data was not available to help explain the issue that was reported from the field.
As a result of this finding, abbott is performing further investigation.
|
|
Search Alerts/Recalls
|