US SURGICAL PUERTO RICO ENDO GRASP; MOTOR, SURGICAL INSTRUMENT, PNEUMATIC POWERED Back to Search Results
Model Number 174233
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted that the plastic cable sheath had shearing damage and was bent.A functional evaluation found that jaw rotation worked without difficulty.Release lever and jaw toggle functioned properly.Ratchet button functioned properly.Jaw articulated, opened and closed without difficulty.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the reported condition may occur where an excessive force or leverage is applied to the device while retracting the clevis assembly and the clevis boot is subsequently damaged.This subsequent damage may result in improper retraction and deployment of the jaws during articulation and may cause shearing of the plastic cable sheath.The root cause of the observed damage was found to be due to the device not being used as intended which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, prior to use on a laparoscopic procedure, when the surgeon was checking extension, rotation and articulation of the device before patient entering the operating room, when it was extended it appeared to shear opaque plastic on the shaft of the instrument.The new instrument was used to complete the case.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GRASP
Common Device Name
MOTOR, SURGICAL INSTRUMENT, PNEUMATIC POWERED
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key8814468
Report Number2647580-2019-03795
Device Sequence Number2657653
Product Code GET
UDI-Device Identifier20884523000815
UDI-Public20884523000815
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K914753
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2023
Device Model Number174233
Device Catalogue Number174233
Device Lot NumberP8J1400X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/19/2019
Initial Report FDA Received Date07/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-