| Model Number |
8886848882 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/04/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during laparoscopic partial nephrectomy, the device did not clamp well, the clamping force was weak.The procedure was completed with another device.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Visual examination of the cartridge noted that the clip was fully formed but not completely deployed from the cartridge.Since the clinical instrument was not received the loading unit was loaded into a pmv representative instrument and fired; the pusher advanced properly and deployed the clip.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the clip not fully deployed condition may occur if the trigger of the instrument is not fully actuated releasing the clip from the cartridge.The root cause of the observed damage was found to be due to the device not being used as intended which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
|
| |
|