• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI ENTERRA; INTESTINAL STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number 4351-35
Medical Device Problem Codes Break (1069); Human-Device Interface Problem (2949)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/12/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a manufacturer representative regarding a patient with an implantable neurostimulator (ins).It was reported that the lead was placed and the surgeon went to secure it with a clip applier, as indicated by protocol, and the clip applier cut the blue polypropylene suture attached to the ski needle.The surgeon removed the lead and replaced it with a new lead without issue.The procedure went well and the patient was doing fine.The issue was resolved at the time of the report.No further complications were reported/anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Common Device Name
INTESTINAL STIMULATOR
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key8816198
Report Number2649622-2019-12835
Device Sequence Number16228874
Product Code LNQ
UDI-Device Identifier00643169936577
UDI-Public00643169936577
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
H990014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/14/2021
Device Model Number4351-35
Device Catalogue Number4351-35
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/12/2019
Initial Report FDA Received Date07/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
-
-