| Catalog Number |
6206000000 |
| Medical Device Problem Code |
Device Remains Activated (1525)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/24/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Reported events 1 event was reported for this quarter. product return status 1 device was received.Event confirmation status 1 reported event was confirmed.Evaluation results 1 device was found to be affected by a damaged drivetrain. additional information 1 device is not labeled for single-use.1 device was not reprocessed and reused.
|
| |
|
Event or Problem Description
|
|
This report summarizes 1 malfunction event in which the device had run-on.1 event had no patient involvement; no patient impact.
|
| |
|
Search Alerts/Recalls
|