Catalog Number RTLR180343
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Patient Involvement (2645)
Event Date 06/04/2019
Event Type  malfunction  
Manufacturer Narrative
Plant investigation: the actual device was returned to the manufacturer for physical evaluation.An exterior visual inspection of the returned cycler showed no signs of physical damage.The reported symptom noisy was not confirmed.The reported fluid temperature out of range alarm was confirmed.The system air leak test passed.The teach pump test passed.The heater calibration test failed.Further investigation led to find a short on f1 of the ac distribution board which was causing the heater tray unable to warm up.An internal visual inspection of the returned cycler encountered no other discrepancies.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.Upon completion of the evaluation, the reported issue was confirmed and the cause was determined to be a short on the distribution board.The cycler was refurbished following the evaluation.
 
Event Description
It was reported that the patient¿s liberty select cycler was noisy and receiving a fluid temperature out of range alarm during step 7 of set up.The patient stated that the loud cracking noise was occurring during the beginning steps of set up.At that point in time, the technical support representative advised the patient to discontinue use of the cycler and to notify their peritoneal dialysis registered nurse (pdrn) of the event.A replacement cycler was issued to the patient.It was reported that an alternate treatment option, manuals was available.Upon follow up, the peritoneal dialysis registered nurse (pdrn) stated that the patient completed treatment manually.There were no adverse events or medical intervention required as a result of the reported event.The cycler was returned to the manufacturer and a replacement cycler was provided and received.Upon physical evaluation of the cycler by the manufacturer, it was identified that there was a short on the ac distribution board.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY SELECT CYCLER ASSY(NON-VALUATED)
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
matthew amaral
920 winter st.
waltham, MA 02451
7816999758
MDR Report Key8816800
MDR Text Key151885948
Report Number2937457-2019-02279
Device Sequence Number1
Product Code FKX
UDI-Device Identifier00840861102068
UDI-Public00840861102068
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181108
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/23/2019
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberRTLR180343
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2019
Device Age MO
Date Manufacturer Received07/15/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/05/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-