This report is submitted beyond 30 calendar days from allergan¿s receipt due to the exemption transition period.A review of the device history record has been completed.No deviations or non-conformance's noted.Device evaluation: analysis of the returned device identified: yellow particles, weigh to the spec, creases flat and deformation device.Bubbles and cloudy, the gel observed after autoclave cycle.Based on the device analysis the final assessment is: no issues found related with the manufacturing process.The event of capsular contracture baker grade iii is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Reason for reoperation: capsular contracture baker grade iii.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
|