| Model Number |
9355-0953-01 |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the associated device displayed a "poor pad contact" message using these electrodes.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The actual electrodes from the event were not available for evaluation.Instead, electrodes from retained samples of the same lot number were investigated.Evaluation of the electrodes did not reveal any irregularities that would have contributed to the reported event.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Additional Manufacturer Narrative
|
|
The electrodes were returned for evaluation; the malfunction was duplicated during visual evaluation.The malfunction was attributed to a disconnected jumper wire (self test wire) on the electrode pad.This type of failure does not disable the electrode from delivering therapy when attached to a defibrillator.This is a malfunction which only impacts self-test of the electrode with the defibrillator.This type of malfunction does not pose any clinical impact and does not meet our guidelines for reportability.Analysis for reports of this type has not identified an increase in trend.
|
| |
|