BOSTON SCIENTIFIC CORPORATION ALTRUA 50; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number S501
Device Problems Premature Discharge of Battery (1057); High Capture Threshold (3266)
Patient Problem Injury (2348)
Event Date 01/26/2019
Event Type  Injury  
Manufacturer Narrative
This device has not been returned; therefore, technical analysis cannot be conducted.Without a returned device it is not possible to definitively confirm how the device may have contributed to the complaint incident.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
It was reported that during a routine follow up appointment, this pacemaker was found to be operating in automatic capture retry mode.An electrocardiogram revealed the patient was in chronic atrial fibrillation (af) with an occasional paced beat at 50 bpm.In august 2018, this pacemaker had an estimated longevity of two years but declared end of life (eol) in may.It is unknown if the rapid battery depletion was due to a lead issue/automatic capture retry or premature battery depletion (pbd).The device was explanted and discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALTRUA 50
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CLONMEL LIMITED
cashel road
,
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
,
saint paul, MN 55112
6515826168
MDR Report Key8818727
MDR Text Key151953278
Report Number2124215-2019-14290
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 07/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/23/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/12/2012
Device Model NumberS501
Device Catalogue NumberS501
Device Lot Number576980
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/09/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/14/2010
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age87 YR
-
-