Model Number 27061
Device Problems Premature Discharge of Battery (1057); Inadequate User Interface (2958)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to resmed and an evaluation confirmed the complaint.The internal battery and top case was replaced to address this issue.The device was serviced and fully tested before it was returned to the customer.Resmed's risk analysis for this failure mode concludes that the risk is acceptable.(b)(4).
 
Event Description
It was reported to resmed that an astral device had an internal battery with a reduced level of capacity and had an unresponsive touchscreen.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 100 - EUR4
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive.
bella vista
sydney, nsw 2153
AU  2153
Manufacturer (Section G)
RESMED LTD
1 elizabeth macarthur drive.
bella vista
sydney, nsw 2153
AU   2153
Manufacturer Contact
michelle hughes
1 elizabeth macarthur drive.
bella vista
sydney, nsw 2153
AU   2153
MDR Report Key8820043
MDR Text Key152043795
Report Number3004604967-2019-00232
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeGB
PMA/PMN Number
K152068
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27061
Device Catalogue Number27061
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/04/2019
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date07/02/2019
Initial Date Manufacturer Received 07/02/2019
Initial Date FDA Received07/23/2019
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-