• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, COSTA RICA LTDA INQUIRY¿ STEERABLE DIAGNOSTIC CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, COSTA RICA LTDA INQUIRY¿ STEERABLE DIAGNOSTIC CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number IBI-81404
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Cardiac Perforation (2513)
Date of Event 07/03/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported pericardial effusion could not be conclusively determined.Per the ifu, vascular perforation is an inherent risk of any electrode placement.
 
Event or Problem Description
Related manufacturing reference: 3008452825-2019-00358, 2182269-2019-00099, 2182269-2019-00100, 2030404-2019-00058.During a pulmonary vein contraction (pvc) ablation procedure a pericardial effusion occurred.Following ablation of the right coronary cusp, the user attempted a second pvc in the left ventricle (lv) through the coronary sinus (cs).A flexibility catheter was used first through an agilis sheath but there was no success in reaching the target site.A safire catheter was then attempted but it was decided the anatomy of the patient was too difficult to place a catheter out distally in the cs.The user decided to forgo further ablation and proceeded with an ep study.During the ep study the patient became hypotensive and echocardiography confirmed fluid in the pericardium necessitating a pericardiocentesis.Once stabilized the patient was transferred out of the operating room.There were no performance issues with any abbott device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INQUIRY¿ STEERABLE DIAGNOSTIC CATHETER
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
MDR Report Key8821471
Report Number3008452825-2019-00359
Device Sequence Number2541099
Product Code DRF
UDI-Device Identifier05414734304146
UDI-Public05414734304146
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K961924
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2022
Device Model NumberIBI-81404
Device Catalogue NumberIBI-81404
Device Lot Number6888618
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/03/2019
Initial Report FDA Received Date07/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AGILIS¿ NXT STEERABLE INTRODUCER; FLEXABILITY¿ ABLATION CATHETER, SENSOR ENABLED; INQUIRY¿ STEERABLE DIAGNOSTIC CATHETER; SAFIRE¿ BI-DIRECTIONAL ABLATION CATHETER
Outcome Attributed to Adverse Event Required Intervention;
Patient Weight109
-
-