• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANIMAS LLC ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Code Image Display Error/Artifact (1304)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event or Problem Description
On (b)(6) 2019, the reporter contacted animas, alleging a display (dim/fading/color spectrum) issue.There is no indication the alleged product issue caused or contributed to an adverse event.This complaint is being reported because the issue may impact the user's ability to read some or all the information on the screen which may result in over or under delivery.
 
Additional Manufacturer Narrative
Device evaluation: the device has been returned and evaluated by product analysis on 02-aug-2019 with the following findings: during investigation, the pump was powered on and the display was found to be dim and discolored.Unrelated to the original complaint, the battery compartment was observed to be cracked.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
MDR Report Key8821891
Report Number2531779-2019-04164
Device Sequence Number2838789
Product Code OYC
UDI-Device Identifier10840406100051
UDI-Public10840406100051
Combination Product (Y/N)Y
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2019
Was the Report Sent to FDA? Yes
Device Age49 MO
Initial Date Received by Manufacturer 07/19/2019
Supplement Date Received by Manufacturer07/19/2019
Initial Report FDA Received Date07/24/2019
Supplement Report FDA Received Date08/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR
Patient Weight228
-
-