| Model Number |
402804 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Material Separation (1562); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/03/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
During a pulmonary vein isolation procedure, the catheter became stuck in the patient.At the start of the procedure there was noise noted on the distal electrodes of the reflexion spiral catheter.The pair were deselected and the procedure was continued.At the end of the case there was resistance when withdrawing the catheter back into the introducer.The curve of the catheter was readjusted but there was no resolution.The introducer and catheter were withdrawn as one unit from the right femoral vein.The catheter was visually inspected and a break in the outside case was noted along with exposure of the push-pull cables at the proximal end of the spiral/shaft connection.Contract injection was used to check for any injury to the patient during catheter/introducer withdrawal with no trauma observed.
|
| |
|
Additional Manufacturer Narrative
|
|
One duo-decapolar, bi-directional, variable radius, reflexion spiral catheter was received for evaluation.The catheter shaft was torn and separated at the spiral loop/shaft transition exposing the internal wires.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event is consistent with damage during use.
|
| |
|
Search Alerts/Recalls
|