| Model Number |
U41502H30 |
| Medical Device Problem Codes |
Material Frayed (1262); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device for this event has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The device is labeled for single use.
|
| |
|
Event or Problem Description
|
|
This report summarizes one malfunction.A review of the reported information indicated that model u41502h30 pta balloon dilatation catheter allegedly had frayed material.This information was received from one source.The malfunction involved a patient without consequence.The age, weight, and sex of the involved patient were not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the lot history review was performed.The device for this event has been returned to the manufacturer for evaluation.The investigation is confirmed for the reported material deformation, as distal tip of the balloon appeared to be wrinkled and deformed.The definitive root cause could not be determined.The device is labeled for single use.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
This report summarizes one malfunction.A review of the reported information indicated that model u41502h30 pta balloon dilatation catheter allegedly had frayed material.This information was received from one source.The malfunction involved a patient without consequence.The age, weight, and sex of the involved patient were not provided.
|
| |
|