Model Number TABLETOP
Device Problems Failure to Power Up (1476); Intermittent Loss of Power (4016)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/08/2019
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: 2019-51232.
 
Event Description
A healthcare professional reported the laser key switch intermittently turned off during a laser procedure.Additional information has been received stating the laser stopped working during a surgical procedure.The laser was unable to be restarted.This occurred at the end of the procedure therefore it was completed with no harm to the patient.
 
Manufacturer Narrative
Additional information is provided.The company service representative examined the system but was unable to replicate the reported event.Unrelated to the reported event, the company service representative observed multiple ¿0¿ error codes in the event log.The company service representative opened the front panel, reseated, secured the keyswitch cable, and insulated contact points to prevent shorting.The system was tested and the laser fired dozens of shots without issue and within energy tolerances.The system was then tested and met all product specifications.The system manufacturing device history record (dhr) was reviewed.Based on qa assessment, the product met specifications at the time of release.The system was found to meet specifications; therefore, the root cause of the reported event cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key8823663
MDR Text Key152855546
Report Number2028159-2019-01369
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP
Device Catalogue Number8065751150
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 07/08/2019
Initial Date FDA Received07/24/2019
Supplement Dates Manufacturer Received09/13/2019
Supplement Dates FDA Received09/16/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-