Model Number AT75186
Medical Device Problem Codes Material Frayed (1262); Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lot number for the malfunction was provided and a lot history review will be performed.The device was returned for evaluation.A visual inspection was performed and frayed fibers were noticed on the balloon.No other anomalies were noted to the device.The balloon was attempted to be inflated, but it was unable to be due to dried blood in the sheath.Therefore, the investigation is confirmed for the identified frayed fibers, as fraying fibers were found on the balloon.The balloon rupture investigation is inconclusive, as the device was unable to be tested.Based upon the available information, the definitive root cause is unknown.The devices are labeled for single use.
 
Event or Problem Description
This report summarizes two (2) malfunction.A review of the reported information indicated that model at75186 pta balloon dilatation catheter allegedly experienced material frayed and a material rupture.This information was received from one source.The malfunctions involved one patient with no known impact to the patient.The (b)(6) year old female patient was (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key8824117
Report Number2020394-2019-01829
Device Sequence Number3258404
Product Code LIT
UDI-Device Identifier00801741062506
UDI-Public(01)00801741062506
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberAT75186
Device Catalogue NumberAT75186
Device Lot NumberGFCY2820
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/30/2019
Initial Report FDA Received Date07/24/2019
Patient Sequence Number1
-
-