• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 018 PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE 018 PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number U8150430
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lot number for the malfunction was provided and a lot history review will be performed.The device has been returned for evaluation; the evaluation identified that the catheter was in returned in two segments.Based upon the available information, the definitive root cause is unknown.The device is labeled for single use.
 
Event or Problem Description
This report summarizes one (1) malfunction.A review of the reported information indicated that model u8150430 pta balloon dilatation catheter allegedly experienced a detached component this information was received from one source.The malfunction involved a patient with no known impact to the patient.The patient¿s age, weight, and gender were not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE 018 PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key8824246
Report Number2020394-2019-01836
Device Sequence Number19542637
Product Code LIT
UDI-Device Identifier00801741065101
UDI-Public(01)00801741065101
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberU8150430
Device Catalogue NumberU8150430
Device Lot NumberCMCY0055
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/30/2019
Initial Report FDA Received Date07/24/2019
Patient Sequence Number1
-
-