• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAS; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723LNAS
Medical Device Problem Code Failure to Cycle (1142)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/17/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call that the insulin pump alarmed compromised force sensor system.The customer¿s blood glucose level was 177 mg/dl at the time of the incident.Customer assisted with troubleshooting and stated that they were able to clear the alarm and able to complete insulin pump rewind.Customer performed displacement test and it was failed.The customer was advised that the insulin pump needed to be replaced and revert to the backup plan.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device alarmed motor error during the basic occlusion test due to loose/protruded drive support disk.Unable to perform basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test, displacement test or verify a33 alarm due to motor error alarm.However, device passed rewind test and displacement test.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAS
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key8825115
Report Number2032227-2019-35658
Device Sequence Number2471296
Product Code OYC
UDI-Device Identifier00643169478602
UDI-Public(01)00643169478602
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723LNAS
Device Catalogue NumberMMT-723LNAS
Device Lot NumberA4723LNASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2019
Initial Date Received by Manufacturer 07/17/2019
Supplement Date Received by Manufacturer08/23/2019
Initial Report FDA Received Date07/25/2019
Supplement Report FDA Received Date09/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age47 YR
Patient Weight200
-
-