MEDTRONIC EUROPE SARL PERCEPTA QUAD CRT-P MRI SURESCAN; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION
|
Back to Search Results |
|
Model Number W4TR04 |
Device Problem
Inappropriate or Unexpected Reset (2959)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/03/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product event summary : the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated a software error.Analysis of the device memory indicated a false alert for lead impedance out of range.Analysis of the device memory indicated that the atrial and ventricular rate histogram data was missing/invalid.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the cardiac resynchronization therapy pacemaker (crt-p) had a partial power on reset, which caused the rate histograms and percentage pacing data to be cleared, but the device continued to operate as programmed and began collecting new data automatically.A device update, delivered via in-clinic interrogation, was recommended by the manufacturer and the device remains in use.It was also reported that the left ventricular lead impedance had one low pacing impedance measurement.It was noted to have been trending low, but stable.The lead remains in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|