| Model Number |
693558 |
| Medical Device Problem Code |
Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) pacing rate was lower than the programmed low rate.It was suspected that this was due to the patient's high potassium levels causing the right ventricular (rv) lead to exhibited t-wave oversensing (twos).The patient's potassium levels were being treated.The crt-d and rv lead remain in use and will be monitored as the patient's potassium levels improve.No patient complications have been reported as a result of this event.
|
| |
|