| Catalog Number |
SSX-285 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
06/28/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation is capsular contracture, baker grade iii.This is a known potential adverse event addressed in the product labeling.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a left side capsular contracture, baker grade iii.Device remains implanted.
|
| |
|
Search Alerts/Recalls
|