MAUDE Adverse Event Report: ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 19AGFN-756
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Death (1802)
Date of Event 06/26/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Further information regarding this event has been requested.The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Event or Problem Description
On (b)(6) 2019, a 21 mm regent heart valve was implanted.Post implant, the leaflets were observed to be immobile.The site attempted to re-position the valve, during which one of the valve leaflets detached.The valve was removed and replaced with a 19 mm mechanical heart valve (model: unknown).The site attempted to re-position the leaflets, but were unable to rotate the device.A post-surgery eco showed that the valve was working normally.The patient expired 48 hours post procedure.Additional information has been requested.
 
Additional Manufacturer Narrative
09/25/2019 mjr ¿ supplemental correction report needed for corrected implanted device serial and model number, d4.09/25/2019 mjr - supplemental report needed to address analysis: h3, h6.An event of inability to rotate the device and patient death two days after procedure was reported.Information from the field indicated that the valve was working normally after the surgery.The results of the investigation are inconclusive since the device remains implanted was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM REGENT HEART VALVE
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
MDR Report Key8828688
Report Number2648612-2019-00052
Device Sequence Number14057850
Product Code LWQ
UDI-Device Identifier05414734005982
UDI-Public05414734005982
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/12/2023
Device Model Number19AGFN-756
Device Catalogue Number19AGFN-756
Device Lot Number6586351
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/27/2019
Supplement Date Received by Manufacturer09/25/2019
Initial Report FDA Received Date07/25/2019
Supplement Report FDA Received Date09/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death; Required Intervention;
-
-