| Model Number |
DL-344-10-K |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/27/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The explanted lead has not yet been returned to neuropace for analysis.However, review of the neurostimulator ecog and lead impedances for the left depth lead were suggestive of a potential lead break.
|
| |
|
Event or Problem Description
|
|
A review of the patient's ecog data indicated lost ecog signal and demonstrated high impedances, suggestive of a possible lead break on the left depth lead.The patient reported hitting her head on her bed, during the winter of 2018 however no other contributing factors were reported by the patient or treating center.The patient underwent a procedure on (b)(6) 2019 to remove the lead and implant a new left depth lead no complications were reported during the procedure and the signals and impedances for the newly implanted lead were within normal limits.
|
| |
|