Model Number 60SF2
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/23/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Visual analysis of the returned device indicates a section of the heat-exchanger broken.The investigation is currently ongoing.Following completion of the investigation, conclusion will be provided.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that during use of the cryocath probe ((b)(4)), on the second freeze it started to profusely leak gas.The device was replaced.No patient injury occurred but the surgery was delayed for several minutes.
 
Additional Manufacturer Narrative
Conclusion: the bellows have been manipulated beyond the design and or material capabilities.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRYOFLEX SURGICAL ABLATION PROBE
Common Device Name
SURGICAL DEVICE, FOR ABLATION OF CARDIAC TISSUE
Manufacturer (Section D)
MEDTRONIC, INC.
3800 annapolis lane
minneapolis MN 55447
MDR Report Key8829937
Report Number3008592544-2019-00031
Device Sequence Number18897754
Product Code OCL
UDI-Device Identifier00763000114619
UDI-Public00763000114619
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K123733
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/21/2021
Device Model Number60SF2
Device Catalogue Number60SF2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2019
Initial Date Received by Manufacturer 06/26/2019
Supplement Date Received by Manufacturer08/22/2019
Initial Report FDA Received Date07/26/2019
Supplement Report FDA Received Date09/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-