DEPUY SPINE INC CONN O/O SD TOP NTCH 5.5X5.5 T; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Catalog Number 179771555
Device Problem Material Split, Cut or Torn (4008)
Patient Problem Not Applicable (3189)
Event Date 05/13/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Visual examination found that the initial threads on the connector were torn.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.With the information provided, a definitive root cause for the torn threads cannot be determined.Noted damage suggests that inadvertently cross threading of the setscrew occurred upon insertion into the tulip head of the connector.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2019, that during the surgery, when the surgeon implanted the rod to the connector with one handed rod approximator and then tighten the connector screw.The surgeon recognized the braking noise, then he checked the condition of the connector and found out that the connector thread was cross threaded.The surgery was completed by using alternative connector.There were no broken fragments generated.There was less than 30 minutes of surgical delay.Procedure was completed successfully.There is no patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONN O/O SD TOP NTCH 5.5X5.5 T
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key8830640
MDR Text Key178378270
Report Number1526439-2019-51887
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034077819
UDI-Public10705034077819
Combination Product (y/n)N
PMA/PMN Number
K160904
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 05/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/26/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number179771555
Device Lot NumberNW207284
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2019
Date Manufacturer Received07/26/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-