• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SROM REAMER BLUNT 15MM STR; HIP INSTRUMENTS : REAMERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SROM REAMER BLUNT 15MM STR; HIP INSTRUMENTS : REAMERS Back to Search Results
Catalog Number 536815
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/28/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.The root cause is attributed to use error.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Territory (b)(6) reports that multiple srom reamers were damaged during a surgery with surgeon.The reamers were used to "ream" through some hardware left in a hip.Since the reamers were used to do this, many of them have been damaged and need to be replaced.No further information is available at this time.Examination of the returned reamer confirmed the complaint; the flutes on the reamer are nicked with some areas missing material.The root cause is attributed to use error.Provided information states the reamers were used to "ream" through some hardware left in a hip.Since the reamers were used to do this, many of them have been damaged and need to be replaced.Reamers are not intended to ream through metal hardware, it's intended purpose is to ream bone.The complaint sample consisted of (1) 536815 srom reamer blunt 15mm straight, lot code so2020001.Review of the as400 found this lot was manufactured in 2015.The investigation could not verify or identify any product contribution to the reported event with the information provided.Based on the investigation, the need for corrective action is not indicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that multiple srom reamers were damaged during a surgery with surgeon.The reamers were used to "ream" through some hardware left in a hip.Since the reamers were used to do this, many of them have been damaged and need to be replaced.No further information is available at this time.No surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SROM REAMER BLUNT 15MM STR
Common Device Name
HIP INSTRUMENTS : REAMERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key8831287
Report Number1818910-2019-99344
Device Sequence Number11360300
Product Code HTO
UDI-Device Identifier10603295172994
UDI-Public10603295172994
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number536815
Device Lot NumberSO2020001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/20/2019
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/23/2019
Initial Report FDA Received Date07/26/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-