|
Investigation evaluation: our laboratory evaluation of the product said to be involved determined that there was a leak from a hole in the middle of the balloon.A functional test was performed on the returned device.A cook dilation syringe (ds-60cc-s) was filled with water and attached to the balloon inflation port.The syringe was placed into an inflation handle, and negative pressure was applied to the balloon.After applying negative pressure, the balloon was inflated with water and the balloon would not hold pressure.A leak was seen from a hole in the middle of the balloon.A visual inspection of the catheter showed no kinks/bends.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.It is unknown if the user applied negative pressure and lubrication to the balloon prior to advancement through the endoscope accessory channel.A possible contributing factor to a leakage is failure to apply negative pressure and lubrication to the balloon prior to advancement.The instructions for use advise the user: "to facilitate passage through the endoscope, apply negative pressure to the catheter." the application of negative pressure will aspirate all residual air from the balloon and ease endoscopic advancement.Negative pressure will also aid in balloon preservation and optimize balloon performance.The instructions for use also advise the user: "maintain balloon deflation with negative pressure and introduce into the accessory channel of the endoscope, advancing in short increments until the dilator is completely visualized endoscopically." the instructions for use advise the user: "apply a lubricating agent to the balloon to facilitate passage through the endoscope accessory channel." this activity will aid in balloon preservation.A possible contributing factor to a leakage in the balloon material is using a compromised inflation device to inflate the balloon.If the pressure reading of the inflation device is inaccurate, this could contribute to over inflation, possibly resulting in damage to the balloon material.A leakage can occur if the balloon comes into contact with a sharp object or a burr in the endoscope channel.Prior to distribution, all hercules 3 stage balloons esophageal are subjected to a leak test to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
|
During an esophageal dilation, the physician used a cook hercules 3 stage balloon esophageal.The balloon was not inflating.There was no reportable information at this time.The device was evaluated on 03-jul-2019 and a leak was seen from a hole in the middle of the balloon.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|