Model Number 15 |
Device Problem
Inappropriate/Inadequate Shock/Stimulation (1574)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/01/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Physio-control evaluated the customer's device and verified the reported issue.Physio replaced the device's therapy pcb assembly to resolve the reported issue.After observing proper device operation through functional and performance testing, the device was returned to the customer for use.
|
|
Event Description
|
The customer contacted physio-control to report that the service indicator is present and an event code is logged in the memory of their device.The event code logged in its memory is indicative of a device failure that could result in a partial loss of defibrillator output energy due to a loss of the negative portion of the biphasic output waveform, resulting in a monophasic shock.There was no report of patient use associated with the reported event.
|
|
Manufacturer Narrative
|
Physio-control further evaluated the removed therapy pcb assembly and determined that the cause of the reported issue was due to a shorted transistor, designator q7.
|
|
Event Description
|
The customer contacted physio-control to report that the service indicator is present and an event code is logged in the memory of their device.The event code logged in its memory is indicative of a device failure that could result in a partial loss of defibrillator output energy due to a loss of the negative portion of the biphasic output waveform, resulting in a monophasic shock.There was no report of patient use associated with the reported event.
|
|
Search Alerts/Recalls
|