| Model Number |
3665 |
| Medical Device Problem Code |
Application Program Freezes or Fails to Launch (4031)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
06/09/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.The results/method and conclusion codes along with investigation results will be provided on the final report.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient could no longer connect with their ipg after an unrelated procedure.As a result, surgical intervention was taken to replace the ipg.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event for inoperable ipg was confirmed.It was determined the device was running the service application software.This condition is consistent with electrocautery use during surgery.Per event details, the ipg no longer communicated following an unrelated surgery.As a result of this finding, actions have been taken to prevent reoccurrence.
|
| |
|