4C8723 |
| Medical Device Problem Codes |
Product Quality Problem (1506); Improper Flow or Infusion (2954)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Lot number - r19c20043 and r19c20142.(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
This report summarizes 1 malfunction events.It was reported that the paper ties of an unspecified number of clearlink system y-type blood/solution sets were loose and fell off the set.This occurred both prior to patient use and during priming.The staff reported that this had caused several issues including: the line dropping and hitting the floor when the packaging was opened (the set had to be replaced), the line becoming tangled, the set being difficult to hang on the iv pole, lines unraveling and hitting the ground after priming, and the drip chambers becoming inverted during setup, leading to air getting into the line during priming.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: two (2) single-use devices were received for evaluation.Visual inspection revealed that the paper ties were loose.Functional testing including pressure testing and clear passage testing was performed for both sets with no issues noted.The reported condition of loose paper ties was verified.The cause of the condition was not determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|