• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US S-ROM*SCREW,6.5MM DIA,25MM LG; BONE SCREWS AND PINS : SCREWS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US S-ROM*SCREW,6.5MM DIA,25MM LG; BONE SCREWS AND PINS : SCREWS Back to Search Results
Catalog Number 556072
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Tissue Damage (2104); No Code Available (3191)
Date of Event 07/09/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The primary surgery was performed on (b)(6) 2008 via tha.It was reported that the revision surgery was performed on (b)(6) 2019 by replacing the cup (manufactured by kyocera, p/n: 121780046), the liner (p/n: 121889146), the stem (p/n: 900533210), the head (p/n: 962700100) due to dissociation of the cup, dislocation caused by the dissociation, breakage of the screw (p/n: 556072).It was confirmed after surgical incision that the screw was broken caused by dissociation, wear of the liner and the trunnionosis, part of the soft tissue discolored to black, metallosis.The surgery was completed without problems, and it was unknown whether there was a surgical delay or not.No further information is available.
 
Additional Manufacturer Narrative
Depuy synthese is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.Product complaint # : (b)(4).Investigation summary: no device associated with this report was received for examination.Depuy synthese considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
BONE SCREWS AND PINS : SCREWS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8833472
Report Number1818910-2019-99417
Device Sequence Number18909952
Product Code LPH
UDI-Device Identifier10603295175926
UDI-Public10603295175926
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/21/2013
Device Catalogue Number556072
Device Lot Number0286072E
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/09/2019
Supplement Date Received by Manufacturer10/01/2019
Initial Report FDA Received Date07/26/2019
Supplement Report FDA Received Date10/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SROM 9/10 18X12X135 30
Outcome Attributed to Adverse Event Required Intervention;
-
-