| Catalog Number |
556072 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Tissue Damage (2104); No Code Available (3191)
|
| Date of Event |
07/09/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
The primary surgery was performed on (b)(6) 2008 via tha.It was reported that the revision surgery was performed on (b)(6) 2019 by replacing the cup (manufactured by kyocera, p/n: 121780046), the liner (p/n: 121889146), the stem (p/n: 900533210), the head (p/n: 962700100) due to dissociation of the cup, dislocation caused by the dissociation, breakage of the screw (p/n: 556072).It was confirmed after surgical incision that the screw was broken caused by dissociation, wear of the liner and the trunnionosis, part of the soft tissue discolored to black, metallosis.The surgery was completed without problems, and it was unknown whether there was a surgical delay or not.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthese is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.Product complaint # : (b)(4).Investigation summary: no device associated with this report was received for examination.Depuy synthese considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|