Model Number 3CX*RX25RW
Medical Device Problem Code Device Damaged Prior to Receipt by User (1570)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/09/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event or Problem Description
The distributor reported to terumo cardiovascular that upon receipt of the shipment there were 48 pieces too few of the device as compared to the ship list.No patient involvement.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations and as indicated by terumo cardiovascular systems in the initial report submitted to the fda.Upon further investigation of the reported event, the following information is new and/or changed: (identification of evaluation codes 4114, 3331, 4246, 4308).Method code #1: 4114 device not returned.Method code #2: 3331 analysis of production records.Results code: 4246 transport/storage problem identified.Conclusions code: 4308 cause traced to transport/storage.The investigation verified that the order was inappropriately picked and shipped.It was determined that was a human error during the picking and shipping process.During that period areas were reinforced to catch up on back order that are experienced at the dc.The root cause was a human error at the distribution center during the picking and shipping process.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPIOX RX25 OXYGENATOR WEST
Common Device Name
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key8833840
Report Number1124841-2019-00205
Device Sequence Number2877036
Product Code DTZ
UDI-Device Identifier00699753450110
UDI-Public(01)00699753450110
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K153213
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date03/31/2022
Device Model Number3CX*RX25RW
Device Catalogue NumberN/A
Device Lot NumberXE22
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/09/2019
Supplement Date Received by Manufacturer08/19/2019
Initial Report FDA Received Date07/26/2019
Supplement Report FDA Received Date08/21/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-