| Model Number |
CNO11 |
| Medical Device Problem Code |
Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
04/26/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The event unit returned for evaluation.A follow-up report will be provided upon completion of investigation.
|
| |
|
Event or Problem Description
|
|
Name of procedure being performed: na (incoming inspection).Detailed description of event: the product did not pass the incoming inspection of po#19-02app because of torn packaging.Details are shown on the attached file.Type of intervention: na.Patient status: no patient involved.
|
| |
|
Event or Problem Description
|
|
Name of procedure being performed: na (incoming inspection).Detailed description of event: the product did not pass the incoming inspection of po#: (b)(4) because of torn packaging.Type of intervention: na.Patient status: no patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was returned to applied medical for evaluation.Upon visual inspection, engineering observed minor creases and scuffs on the c-film side of the pouch.Engineering also confirmed that there was a hole in the tyvek of the pouch.Based on the condition of the returned unit, the reported event was caused by the interaction between the tray edge and the pouch.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
|
| |
|