U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA CANADA CORP. MITROFLOW LXA; TISSUE HEART VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA CANADA CORP. MITROFLOW LXA; TISSUE HEART VALVE Back to Search Results
Model Number LXA19
Medical Device Problem Codes Calcified (1077); Gradient Increase (1270); Material Too Rigid or Stiff (1544)
Health Effect - Clinical Codes Aortic Insufficiency (1715); Aortic Valve Stenosis (1717); Calcium Deposits/Calcification (1758); Death (1802)
Date of Event 03/29/2017
Type of Reportable Event Death
Event or Problem Description
A patient was implanted with a mitroflow lx19 (sn (b)(4)) on (b)(6) 2013.The patient death was reported to be related to the premature failure of the mitroflow valve.Medical records reported from (b)(6) medical center reported that the valve failure was due severe valve calcification and leaflets fixed.On (b)(6) 2017 toe reported that ¿ the aortic valve leaflets are severely thickened with reduced systolic excursion.Severe aortic calcification is visualized.Mean gradient was reported to be 97mmhg.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITROFLOW LXA
Common Device Name
TISSUE HEART VALVE
Manufacturer (Section D)
LIVANOVA CANADA CORP.
5005 north fraser way
burnaby, bc V5J 5 M1
CA  V5J 5M1
MDR Report Key8834868
Report Number1718850-2019-01045
Device Sequence Number2078720
Product Code LWR
UDI-Device Identifier00896208000092
UDI-Public(01)00896208000092(240)LXA19(17)171031
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Reporter Type Importer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/03/2019
Report Date (Section F) 07/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2017
Device Model NumberLXA19
Device Catalogue NumberLXA19
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Distributor Facility Aware Date07/03/2019
Device Age55 MO
Event Location Hospital
Date Report to Manufacturer07/03/2019
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date07/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death; Hospitalization;
Patient Age69 YR
-
-