• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER MEDICAL-KALAMAZOO PERFORMANCE PRO AMBULANCE COT; STRETCHER, WHEELED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL-KALAMAZOO PERFORMANCE PRO AMBULANCE COT; STRETCHER, WHEELED Back to Search Results
Catalog Number 6086000000
Medical Device Problem Code Difficult or Delayed Positioning (1157)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 04/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr is part of the fda voluntary malfunction summary reporting program.The device was evaluated in the field and the issue was confirmed; there was a damaged/broken component.The device was repaired and returned.There was no remedial action taken.This device is not labeled for single use.
 
Event or Problem Description
This report summarizes 1 malfunction event, where it was reported it was difficult to load/unload the cot in the ambulance.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORMANCE PRO AMBULANCE COT
Common Device Name
STRETCHER, WHEELED
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
MDR Report Key8835825
Report Number0001831750-2019-00695
Device Sequence Number2514036
Product Code FPO
UDI-Device Identifier07613327261554
UDI-Public07613327261554
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number6086000000
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/01/2019
Initial Report FDA Received Date07/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-