Catalog Number 3005S3PX3
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 04/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.5 devices were evaluated in the field and the issue was confirmed; 4 devices had calibration issues and 1 had a broken/damaged component.The devices were repaired and returned.1 device was evaluated in the field and the issue was confirmed.However, there was no product malfunction; the issue was the result of use error as the scale was not properly zeroed before use.There was no remedial action taken.This device is not labeled for single use.
 
Event or Problem Description
This report summarizes 6 malfunction events, where it was reported the scale was inaccurate.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S3 PX3-3005
Common Device Name
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
MDR Report Key8835849
Report Number0001831750-2019-00656
Device Sequence Number14100086
Product Code FNL
UDI-Device Identifier07613327278354
UDI-Public07613327278354
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized6
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number3005S3PX3
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/01/2019
Initial Report FDA Received Date07/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-