| Model Number |
ME1224 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The sets were received as an unexpected return with an unknown issue although the customer stated they have samples that are leaking from previous complaints.Testing observed a leak at the engagement between each set's tubing and 4-way parallel connector components.Further inspection under magnification noted no issues with any of the tubing depths within the parallel connector.The root cause was identified as a sink condition in the parallel connector component (supplier issue).
|
| |
|
Event or Problem Description
|
|
The set was received as an unexpected return with an unknown issue.Although the customer stated ; "we have samples to send that are leaking from a previous complaints it is unknown what the product failure is on the unexpected set.
|
| |
|