LIVANOVA CANADA CORP. MITROFLOW LXA; TISSUE HEART VALVE Back to Search Results
Model Number LXA19
Medical Device Problem Codes Calcified (1077); Degraded (1153); Gradient Increase (1270); Material Too Rigid or Stiff (1544)
Health Effect - Clinical Codes Aortic Insufficiency (1715); Aortic Valve Stenosis (1717); Calcium Deposits/Calcification (1758); Death (1802)
Date of Event 03/29/2017
Type of Reportable Event Death
Additional Manufacturer Narrative
The manufacturing and material records for the mitroflow bioprosthetic pericardial heart valve, model # lxa19, s/n # (b)(4), as they pertain to the reported event, were retrieved and reviewed by quality engineering at livanova (b)(4) corp.The results confirmed that this valve satisfied all material, visual, and performance standards required for a lxa19 mitroflow aortic pericardial heart valve at the time of manufacture and release.The device is not available for return and no further information is available for the reported event.The manufacturer is unable to perform any further investigations.Structural valve deterioration is a known inherent risk of failure for biological implants.However, because no further investigations are possible the root cause cannot be established at this time.
 
Event or Problem Description
A patient was implanted with a mitroflow lx19 (sn (b)(4)) on (b)(6) 2013.The patient death was reported to be related to the premature failure of the mitroflow valve.Medical records reported from (b)(6) medical center reported that the valve failure was due severe valve calcification and leaflets fixed.On (b)(6) 2017 toe reported that ¿ the aortic valve leaflets are severely thickened with reduced systolic excursion.Severe aortic calcification is visualized.Mean gradient was reported to be 97mmhg.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITROFLOW LXA
Common Device Name
TISSUE HEART VALVE
Manufacturer (Section D)
LIVANOVA CANADA CORP.
5005 north fraser way
burnaby, bc V5J 5 M1
CA  V5J 5M1
MDR Report Key8837280
Report Number3004478276-2019-00230
Device Sequence Number16693494
Product Code LWR
UDI-Device Identifier00896208000092
UDI-Public(01)00896208000092(240)LXA19(17)171031
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P060038
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2017
Device Model NumberLXA19
Device Catalogue NumberLXA19
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/03/2019
Initial Report FDA Received Date07/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death; Hospitalization;
Patient Age69 YR
-
-