Ppf alleges stroke/heart attack, bone fracture and infection requiring iv antibiotics.At this time it¿s unknown where the fracture occurred.If/when more information is received, the pc will be updated as needed.Ppf also indicates that the patient is deceased.Date and reason of death were not provided.Doi: (b)(6) 2008.Dor: (b)(6) 2008, (right hip).
|
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot number was not provided.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|