| Model Number |
314-13-23 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Inadequate Osseointegration (2646)
|
| Date of Event |
06/07/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Pending evaluation.
|
| |
|
Event or Problem Description
|
|
Primary surgery: (b)(6) 2019.Surgeon had to do surgery on a patient with nonunion of the tub.Minus on left shoulder.He treated the nonunion without complications.While he was doing so, he realized that the poly of the glenoid was loose.The cage and two pegs where tightly connected to the bone; one peg was still connected to the poly but not integrated in the bone.The poly was sheered of the cage and the two integrated pegs.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h10: (d4) serial number: (b)(6), expiration date: 18-mar-2020.(d8) is this a single-use device that was reprocessed and reused on a patient?: no.(e3) occupation: physician.(g5) pma/510(k)number: k113309.(h3) the revision reported was likely the result nonunion.The reported loosening may have either been the result of drilling the peripheral peg holes at an angle or not fully seating the cage glenoid component at the time of implantation, which led to the disassociation of the metal pegs from the polyethylene body of the glenoid component.However, this cannot be confirmed because the component was not returned for evaluation and x-rays were not provided.(h4) device manufacture date: 25-mar-2015.(h6) evaluation codes: 1924, 4002 section h11: the following sections have corrected information: (h5) labeled for single use?: yes.
|
| |
|