| Model Number |
7010 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/28/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Returned product consisted of a quantum maverick balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 58.7cm distal of the strain relief.There was blood in the guidewire lumen.The balloon was tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 08jul2019.It was reported that shaft kink occurred.A 3.25mm x 12mm quantum maverick balloon catheter was advanced for dilatation.However, it was noted that the delivery sheath was kinked.The procedure was completed with another of the same device.There were no patient complications nor injuries reported and the patient's status was stable.However, returned device analysis revealed shaft was separated.
|
| |
|