| Model Number |
7005 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/28/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 52.7cm distal of the strain relief.There was contrast and blood in the inflation lumen, contrast in the balloon, and blood in the guidewire lumen.The balloon was loosely folded.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 08jul2019.It was reported that shaft kinked occurred.The 85% stenosed, 30-33mmx4.0-5.0mm target lesion was located in the moderately tortuous and mildly calcified left anterior descending artery.A 4.0mm x 8mm quantum maverick ballon catheter was advanced for treatment.However, it was noted that the balloon delivery shaft was kinked.The procedure was completed with another of the same device.No patient complications reported and the patient's status was stable.However, returned device analysis revealed shaft detached/separated.
|
| |
|