Model Number 7005
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/28/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 52.7cm distal of the strain relief.There was contrast and blood in the inflation lumen, contrast in the balloon, and blood in the guidewire lumen.The balloon was loosely folded.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event or Problem Description
Reportable based on device analysis completed on 08jul2019.It was reported that shaft kinked occurred.The 85% stenosed, 30-33mmx4.0-5.0mm target lesion was located in the moderately tortuous and mildly calcified left anterior descending artery.A 4.0mm x 8mm quantum maverick ballon catheter was advanced for treatment.However, it was noted that the balloon delivery shaft was kinked.The procedure was completed with another of the same device.No patient complications reported and the patient's status was stable.However, returned device analysis revealed shaft detached/separated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8837852
Report Number2134265-2019-08912
Device Sequence Number14050748
Product Code LOX
UDI-Device Identifier08714729392606
UDI-Public08714729392606
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/07/2021
Device Model Number7005
Device Catalogue Number7005
Device Lot Number0022912599
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2019
Initial Report FDA Received Date07/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age68 YR
Patient Weight62
-
-