• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ULTRA FAST-FIX AB ASSEMBLY - REV CURVE; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ULTRA FAST-FIX AB ASSEMBLY - REV CURVE; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE Back to Search Results
Catalog Number 72201495
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/04/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
It was reported that during the surgery, t2 did not trigger.A backup device was used to complete the surgery.No delay or patient injury reported.
 
Additional Manufacturer Narrative
One curved ultra fast-fix ab assembly for use in treatment, has not been returned for evaluation.Without the reported product a visual and functional evaluation cannot be performed and the customer¿s complaint cannot be confirmed.An exact root cause cannot be determined with confidence.From the information provided, t2 would not deploy.However, factors that could have contributed to the reported event include: bending of the delivery needle.Pathological conditions in the soft tissue that would prevent secure fixation.The instruction for use was reviewed and was found to outline precautionary statements and instructions in regards to the use of the device to avoid damage or non-functionality.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRA FAST-FIX AB ASSEMBLY - REV CURVE
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key8838433
Report Number1219602-2019-00886
Device Sequence Number14099940
Product Code GAT
UDI-Device Identifier03596010597236
UDI-Public03596010597236
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K151105
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/04/2024
Device Catalogue Number72201495
Device Lot Number50791023
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/08/2019
Supplement Date Received by Manufacturer08/29/2019
Initial Report FDA Received Date07/29/2019
Supplement Report FDA Received Date08/30/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-