| Catalog Number |
406047 |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that an unspecified number of tray epid cont we18g3.5 c20 lor5 s/l/l-e experienced a mix of product types in a pack which was noted prior to use.The following information was provided by the initial reporter: material no: 406047 batch no: 0001285103.Voicemail received which mentioned an issue with item 406047.The voicemail cut in an out but it seems that one of the components in the tray is not correct.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: dhr: a review of the device history records identified no issues for lot 0001285103.No samples were returned for analysis.The investigation noted that packaging of the 406047 is a manual process.As part of this manufacturing process, all components are to be verified with the applicable bill of material at start up and at any time during the manufacturing process when more components are delivered to the manufacturing line.The investigation identified the most probable root cause to be a failure in the component verification step with the applicable bill of materials during the manufacture of the affected finished good lot.
|
| |
|
Event or Problem Description
|
|
It was reported that an unspecified number of tray epid cont we18g3.5 c20 lor5 s/l/l-e experienced a mix of product types in a pack which was noted prior to use.The following information was provided by the initial reporter: material no: 406047 batch no: 0001285103 voicemail received which mentioned an issue with item 406047.The voicemail cut in an out but it seems that one of the components in the tray is not correct.
|
| |
|
Search Alerts/Recalls
|