| Catalog Number |
SCM-255 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
04/11/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This report is submitted beyond 30-calendar days from allergan¿s receipt due to the exemption transition period.A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified flat creases and deformation.Cloudy gel was observed after autoclave cycle.A weight test of the explanted material was verified and it was within specification.Based on the device analysis the final assessment is: no observations found related with the complaint reported by the physician.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.The reason for reoperation is capsular contracture, baker grade iv.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.This is a known potential adverse event addressed in the product labeling.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side capsular contracture, baker grade iv.Device has been explanted.
|
| |
|